Using the Monitor
SOLUTION:
Check that calculations have been corrected
appropriately for the relevant variable. In general,
calculated saturation values are not as reliable as direct
laboratory hemoximeter measurements.
If there is excessive light, cover the sensor with opaque
material.
Circulation distal to the sensor site should be checked
routinely. The site must be inspected every 8 hours to
ensure adhesion, skin integrity, and correct optical
alignment. If skin integrity changes, move the sensor to
another site.
Try to keep the patient still, or change the sensor site to
one with less motion.
Observe all instructions, warnings, and cautions in this
manual and in the directions for use of the sensor.
PROBLEM: A valid SpO
2
signal was present but has
disappeared.
CAUSE:
A BP determination on the same limb is in progress.
SOLUTION:
An alarm message codes will appear on the screen, and
the audible alarm will sound immediately.
PROBLEM: An SpO2 PLACEMENT error has occurred.
CAUSE:
Weak or “noisy” signal.
SOLUTION:
Check the patient.
If possible, keep the patient still; check whether the
sensor is applied securely and properly, and replace it if
necessary; move the sensor to a new site; or replace the
sensor.
Change sensor type.
Consider increasing perfusion using heat.