60
Note: If the SpO
2
plethysmograph is displayed on the LCD,
the waveform pauses for 2 minutes or until the Clear button
is selected. SpO
2
values are also retained the same manner
as the BP and Temperature values.
Print
Selection of this button causes the current data to be printed.
Notes
• The Print button appears only when Print is set to
Manual mode.
• If the printer is in Auto print mode, the data will be
printed automatically.
More... Button
Selection of this button displays the More... menu. The
More... menu has six options (depending on model of
Monitor), most of which have submenus. For this reason,
instructions for the More... button are in a separate section.
Set BP Button (UK: BP Mode)
Selection of this button displays the Auto, Stat, and Manual
BP menu.
Auto
Selection of this option starts an Auto BP determination.
When Auto Mode is selected, a number at the right of the
Auto button indicates the time interval between each
reading. To change the time interval, choose the box around
the number and turn the rotor until the desired interval is
reached. The interval can be set between 1 and 120 minutes
(1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60, 90, and 120 minutes).
Press the rotor to confirm the setting. After power-off, the