pulsation and inaccurate measurement of oxygen
saturation.
Excessive pressure from the sensor may cause necrosis
of the skin.
For additional warnings and information, refer to the
NELLCOR
®
sensor’s directions for use.
Monitor performance:
•When an SpO
2
sensor is on a limb that has a blood
pressure cuff, the SpO
2
data will not be valid when the
cuff is inflated. If SpO
2
readings are required during
the entire blood pressure determination, attach the
SpO
2
sensor to the limb opposite the one with the
blood pressure cuff.
Remove nail polish and artificial nails. Placing a sensor
on a polished or an artificial nail may affect accuracy.
Cautions
Patient safety:
Do not place any clip-on sensor in a patient’s mouth or
on a patient’s nose or toe.
Do not place a clip-on finger sensor on a patient’s
thumb or across a child’s foot or hand.
The sensor disconnect error message and associated
alarm indicate that the sensor is either disconnected or
the wiring is faulty. The user should check the sensor
connection and, if necessary, replace the sensor, cable,
or both.
Observe the sensor site to assure adequate distal
circulation. Sensor sites should be checked at least
every 2 hours and rotated at least every 4 hours.
Monitor performance:
Placing a sensor distal to an arterial line may interfere
with adequate arterial pulsation and compromise the
measurement of SpO
2
.
Place the sensor so that the LEDs and the photodiode
are opposite each other.