Procedural and Error Alarm Codes (cont.)
Alarm
Code
LED
Display
LCD Description
Audible Tone
and Volume
Effect of Alarm
Silence Switch
Effect of
Clear via
SelectKnob
Probable Cause
P55
No
change
P55 -
SpO
2
NO SIGNAL
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear
No or very low SpO
2
signal.
Check or reposition sensor
P00
No
change
P00 -
SpO
2
SENSOR
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear
SpO
2
sensor not connected.
No sensor code detected.
Sensor failure
No
Code
No
change
SpO
2
PLACEMENT?
(Masimo Only)
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear
SpO
2
signal weak or noisy.
Sensor failure
No
Code
Values
zeroed
SpO
2
SENSOR OFF?
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear SpO
2
Sensor off patient
E33
No
change
E33 -
TEMP: FAIL
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear
Temperature probe not
connected or inoperable
E00
No
change
E00 -
TEMP: FAIL
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear
Predictive temperature
determination > 60 sec or
attempting axillary temp
No
Code
Blank HIGH TEMP
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear
Predictive temperature exceeds
upper range
No
Code
No
change
TEMP PROBE
BROKEN
High priority
alarm. Volume
adjustable
2 minutes
silence
Clear Broken temperature probe