Revision A PRO 100-400 V2 Patient Monitor Service Manual 2-13
2009829-001
Product Description: SECTION 2. PRODUCT DESCRIPTION
2.6.3. Mechanical
Dimensions Height: 9.8 in (25.0 cm)
Width: 9.8 in (25.0 cm)
Depth: 6.9 in (17.5 cm)
Weight, Including Battery 7.8 lb (3.5 kg)
Mountings Self-supporting on rubber feet or pole mountable
Classification Information Mode of operation: continuous Degree of protection against
harmful ingress of water: Drip-proof IPX1
2.6.4. BP
Cuff Pressure Range
(Normal operating range)
0 to 290 mmHg (adult)
0 to 140 mmHg (neonate)
Default Target: Cuff Inflation 160 ± 15 mmHg (adult)
110 ± 15 mmHg (neonate)
Target Cuff Inflation:
Adjustment Range (5 mmHg increments)
100 to 250 mmHg (adult)
100 to 140 mmHg (neonate)
Blood Pressure Measurement
Adult Range (mmHg)
Systolic 30-245
MAP 15-215
Diastolic 10-195
Neonate
Systolic 40-140
MAP 30-115
Diastolic 20-100
Blood Pressure Accuracy
Meets or exceeds ANSI/AAMI standard SP-10 (mean error
≤5
mmHg, standard deviation
≤8 mmHg)
Maximum Determination Time 120 seconds (adult)
85 seconds (neonate)
Overpressure Cutoff 300 to 330 mmHg (adult)
150 to 165 mmHg (neonate)
Pulse Rate Range 30 to 200 beats/min (adult)
30 to 200 beats/min (neonate)
Pulse Rate Accuracy ±3.5%
US Patents: 4,360,029; 4,501,280; 4,546,775; 4,638,810; 5,052,397; 4,349,034; 4,543,962; 4,627,440; 4,754,761; 5,170,795
European Patents EP122123, EP205805, EP207807