3-6 PRO 100-400 V2 Patient Monitor Service Manual Revision A
2009829-001
Theory of Operation: 3.2 OVERALL PRINCIPLES OF OPERATION
Main Board calculates the pulse rate/Mean Arterial Pressure (MAP).
The results are then displayed on the front panel (seven segment
Light Emitting Diodes (LEDs) and sent to the printer.
The Pneumatic Valve/Manifold (PVM) device has an overpressure
sensor built-in to protect against over-inflation. If an over-inflation
condition occurs, the OVERPRESSURE signal is routed to the PVM
to release the air pressure. The Main Board also, generates an alarm
condition with the speaker sounding and a message in the Liquid
Crystal Display (LCD).
3.2.2 Temperature (Model 200 and 400)
The PRO Monitor uses Alaris Turbo Temp technology to measure
patient temperature. The Turbo Temp probe contains a heating
element that pre-heats the probe to reduce determination time. The
heating function is controlled by the Main Board. The Turbo Temp
probe also contains a thermistor that indicates the temperature.
When the probe is attached to the temperature connector and patient,
the signal generated by the thermistor is routed to the Main Board.
The Main Board converts the thermistor signal along with status
information (i.e ORAL or RECTAL probe indicators) to a DIGITAL
signal. The Main Board then processes the DIGITAL signal and
displays the patient temperature on the UI Board and printer in
Celsius or Fahrenheit.
3.2.3 SPO2 (Model 300 and 400)
The SPO2 probe has a built-in sensor. When the SPO2 sensor is
attached to the SPO2 connector and patient, the probe senses the
heart rate and oxygen saturation. These analog signals are routed to
the SPO2 PWA. The analog signals are analyzed on the SPO2 PWA.
The results are digitized and sent to the Main Board via opto
couplers. The couplers provide for patient isolation as well as serial
data interface. The Main Board processes the data and routes it to
the appropriate displays and/or printer.
A reset signal to the SPO2 PWA is also provided so that the power up
sequencing is corrected. If the SPO2 circuit quits communicating to
the Main Board, the Main Board will attempt to reset the SPO2 PWA.
3.2.4 Host Communication Port
The Host Comm Port is used to interface the Monitor with other
electronic devices (a central nurse’s station or remote alarm device.)
Signals can be sent to the Monitor to initiate blood pressure
determinations and other functions. Patient data can also be retrieved
through this port. For further information, reference the DINAMAP
PRO 100-400 Series Host Communication manual.