Revision A PRO 100-400 V2 Patient Monitor Service Manual 4-9
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
4.4.2. Clinician Mode Main Menu
The Clinician Mode Main menu, pictured above, provides access to
the options normally required for routine use of the PRO Monitor. The
individual buttons on this menu are described briefly below.
Press Button
Selecting the Press button displays the target pressure dialog box,
pictured above, which sets the default target inflation pressure for a
BP cycle. The factory default is 160 mmHg for adults and 110 mmHg
for neonates. This is indicated by the “Auto” label at either end of the
adjustable range. Choosing “Retain Values” allows the user to preset
the number of minutes for which displayed values are retained on the
Monitor.
Selecting the Info button displays the most recent calibration dates of
the BP. Choose OK to return to the Service menu.
Selecting the Silence button displays the dialog box, pictured above,
which mutes all the alarms except the fail-safe alarm. The alarms are
disabled until the Monitor is either powered off and on again or the