Revision A PRO 100-400 V2 Patient Monitor Service Manual 2-15
2009829-001
Product Description: SECTION 2. PRODUCT DESCRIPTION
Low Perfusion 20 to 250 beats/min ±3 digits
* Adult specifications are shown for OXIMAX MAX-A and MAX-N sensors. Neonate specifications are shown for OXIMAX MAX-N. Saturation
accuracy may vary by sensor type.
** Applicability: OXIMAX MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N sensors.
Default Settings
SpO
2
(%) HIGH: 100
SpO
2
(%) LOW: 90
Response Mode 2 (for Mode 2: Fast Response)
Sat Seconds 0
Audible indicator Pitch changes continuously with saturation; volume from 0
(off) to 9
Waveforms Pulse plethsysmograph waveform on LCD gain compensated
Sensor Connect/Disconnect From Patient Monitor detect attachment or disconnection of sensor from
patient within 15 seconds.
Pulse Detection Monitor will detect pulse or enter no signal state within 15
seconds of being attached to patient
Loss of Pulse Monitor will detect loss of pulse from patient and enter no
signal state within 10 seconds
Sensor Light Source
Wavelength Infrared: 890 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation 52.5 mW (max)
Nellcor Patents
4,621,643; 4,653,498; 4,700,708; 4,770,179; 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,421,329;
5,845,847; 5,533,507; 5,577,500; 5,803,910; 5,853,364; 5,865,736; 6,083,172; Re. 35,122 and, international equivalents.
Masimo SET SpO
2
Measurement Range
SpO
2
1 to 100%
Pulse Rate 25 to 240 beats/min
Perfusion Range 0.02 to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion - Adult/Ped* 70 to 100% ±2 digits
Without Motion - Neonate* 70 to 100% ±3 digits
With Motion - Adult/Ped/Neo**† 70 to 100% ±3 digits