4-12 PRO 100-400 V2 Patient Monitor Service Manual Revision A
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
2. Select the Service button.
3. The service menu requires an access code. Turn the rotor to
highlight a number, and push to make a selection. Enter access code
2 2 1 3.
4.5.2 Service Mode Main Menu
The options accessed through the Main Service menu, pictured
above, allow the technician to check basic functions of the PRO
Monitor, calibrate the NIBP system, and set the language of
operation.
Note: These menus and adjustments should be used only by
qualified service technicians.