Revision A PRO 100-400 V2 Patient Monitor Service Manual 4-23
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
Figure 4-2
Adjusting the overpressure requires opening the PRO Monitor and
adjusting the overpressure potentiometers on the Main board.
1. Enable Service Mode.
a. Power on the Monitor.
b. Select the More… button from the Main menu.
c. Select the Service button.
d. Enter the Service Mode access code 2 2 1 3.
e. Select NIBP.
2. Adjust potentiometer VR1 fully counter-clockwise.
3. Manually inflate the system pressure to 315mmHg (the optimal
default Adult Overpressure setting.)
4. Once the manometer displays 315 mmHg, adjust VR1 clockwise
until the valves release the pressurized system.
5. Perform Check again, verifying that the overpressure trips at 315
mmHg ±10 mmHg.
6. Power the Monitor off.
7. Enable Service Mode.
a. Power on the Monitor.
b. Select the More… button from the Main menu.
c. Select the Service button.